Digital Microwave Extraction System , 55ml 8pcs Microwave Assisted Acid Digestion
Product Introduction
YMW series closed microwave digestion system provides a fast, safe and automated solution for samples. It speeds up the sample digestion reaction under high pressure conditions. It is widely used in food, environmental protection, disease control, quality supervision, and Commodity inspection, scientific research institutes and other fields.
Applications
- Food and medicine fields
- Environmental animals and plants fields
- Material metallurgy fields
Product Display

Product Parameters
- Microwave source: 2450MHz, high-energy microwave field emission, dual magnetron emission;
- Machine installation power: 2500W;
- Maximum microwave output power: 2300W, continuous automatic microwave control;
- 7"LCD color touch-screen real time display the temperature and pressure curve;
- Microwave oven volume: 67L large volume and 316 thick stainless steel cavity, multi-layer corrosion-resistant PFA Teflon coating;
- Rotors design: Coaxial one-way rotation, can be adapted to a variety of tank rotors;
- Safety explosion-proof door design: Double-layer metal structure safety door, electronic and mechanical dual control;
- Safety: Intelligent judgment of abnormal state inside the cavity (abnormal sound check, abnormal temperature detection, all electronic equipment operation check) door state monitoring, microwave start is prohibited if the door is not closed;
- Working environment temperature/humidity: Working temperature: 0-40℃, working humidity: 15-85%RH;
- Pressure measurement system: Piezoelectric crystal pressure sensor, measuring range: 0-15MPa (2200psi), accuracy ±0.001MPa;
- Temperature measurement control system: Standard high-precision platinum resistance, temperature measurement range: 0-350℃, accuracy ±0.5℃, temperature control accuracy ±1℃ (optional high-precision multi-core optical fiber temperature sensor, temperature measurement range: 0-300 ℃, accuracy ±0.1℃, control accuracy ±0.1℃;
- Temperature monitoring of the whole reaction tank (optional): Optional IR infrared whole tank scanning temperature measurement system, temperature measurement range: 0-350℃, accuracy ±1℃;
- Furnace cavity exhaust system: High-power corrosion-resistant axial flow fan, turbulent university air cooling, only 20 minutes from 200℃ to 70℃;
- Machine dimension: 630×530×630mm
Model No. | YMW-80 | YMW-100S |
Vessel Type | 1-14pcs closed vessel | 20pcs closed vessel |
Inner Tank Material | TFM | TFM |
Volume Per Vessel | 80ml | 100ml |
Maximum Pressure | 14MPa(2000psi) | 6MPa (850psi) |
Max.Temperature | 250℃ | 250℃ |
Max. Capacity Per Vessel | 0.1-0.5g | 0.1-0.5g |
HF Proof | YES | YES |
Tank Anti-explosion | Material distort pressure release,resume automatically, no need consumables | Imported aluminium alloy anti-explosion film |
Product Characters
- New design uniform heating
- large oven cavity
- High-precision temperature measurement system
- Frequency conversion temperature control system
- Ultimate security system
- Product accessories on demand